Revision
1.0
Last Updated
2025/09/17
Problem Error
Read error: read /dev/usbtmc0: connection timed out
Explanation of Problem
The PC is unable to talk to the analyzer over the USB interface. Most likely the software cannot talk to the analyzer due to a locked communication that has not cleared.
Solution 1 - Clear the analyzer lock
- Check that USB cable is connected to both NUC and analyzer.
- Press power button (red) to turn off the analyzer.
- Press power button (red) to turn on the analyzer.
Solution 2 - Reset Everything
- Perform solution 1 steps 1 and 2, then continue down here.
- Unplug the power from the analyzer, wait 30 seconds or until the power button is no longer red.
- Restart the PC.
- Click on the power button on the top right of the screen
- Select the reboot option
- While the PC is rebooting plug power to analyzer and turn on
- When PC restarts open Firefox and initiate a new test.
- Note if the error repeats or test passes.