Revision

1.0


Last Updated


2025/09/17


Problem Error

Read error: read /dev/usbtmc0: connection timed out



Explanation of Problem


The PC is unable to talk to the analyzer over the USB interface. Most likely the software cannot talk to the analyzer due to a locked communication that has not cleared.

Solution 1 - Clear the analyzer lock

  1. Check that USB cable is connected to both NUC and analyzer.
  2. Press power button (red) to turn off the analyzer.
  3. Press power button (red) to turn on the analyzer.

Solution 2 - Reset Everything

  1. Perform solution 1 steps 1 and 2, then continue down here.
  2. Unplug the power from the analyzer, wait 30 seconds or until the power button is no longer red.
  3. Restart the PC.
    1. Click on the power button on the top right of the screen
    2. Select the reboot option
  4. While the PC is rebooting plug power to analyzer and turn on
  5. When PC restarts open Firefox and initiate a new test.
  6. Note if the error repeats or test passes.